Measurement
Verification of electrical and non-electrical quantities
Counters of what quantities we are able to verify in our laboratories.
Electrical quantities
Accuracy given in ± (% of reading +% of range)
DC voltage
range |
accuracy 1 year after calibratio |
100 mV |
0.0050 + 0.0035 |
1 V |
0.0040 + 0.0007 |
10 V |
0.0035 + 0.0005 |
100 V |
0.0045 + 0.0006 |
1000 V |
0.0045 + 0.0010 |
Alternating current – True RMS ranges 100 mV, 1 V, 10 V, 100 V, 750 V
frequency |
accuracy 1 year after calibration |
3 – 5 Hz |
1.00 + 0.03 |
5 – 10 Hz |
0.35 + 0.03 |
10 Hz – 20 kHz |
0.06 + 0.03 |
20 – 50 kHz |
0.12 + 0.05 |
50 – 100 kHz |
0.60 + 0.08 |
100 – 300 kHz |
4.00 + 0.50 |
Electrical resistance
range |
accuracy 1 year after calibration |
100 Ω |
0.010 + 0.004 |
1 kΩ |
0.010 + 0.001 |
10 kΩ |
0.010 + 0.001 |
100 kΩ |
0.010 + 0.001 |
1 MΩ |
0.010 + 0.001 |
10 MΩ |
0.040 + 0.001 |
100 MΩ |
0.800 + 0.010 |
Direct current
range |
accuracy 1 year after calibration |
100 µA |
0.050 + 0.025 |
1 mA |
0.050 + 0.006 |
10 mA |
0.050 + 0.020 |
100 mA |
0.050 + 0.005 |
1 A |
0.100 + 0.010 |
3 A |
0.200 + 0.020 |
10 A |
0.120 + 0.010 |
Alternating current rozsahy 100 µA, 1 mA, 10 mA, 100 mA
frequency |
accuracy 1 year after calibration |
3 Hz – 5 kHz |
0.10 + 0.04 |
5 – 10 kHz |
0.10 + 0.04 |
range1 A
frequency |
accuracy 1 year after calibration |
3 Hz – 5 kHz |
0.10 + 0.04 |
5 – 10 kHz |
0.10 + 0.04 |
range 3 A
frequency |
accuracy 1 year after calibration |
3 Hz – 5 kHz |
0.23 + 0.04 |
5 – 10 kHz |
0.23 + 0.04 |
range 10 A
frequency |
accuracy 1 year after calibration |
3 Hz – 5 kHz |
0.15 + 0.04 |
5 – 10 kHz |
0.15 + 0.04 |
Frequency **ranges of voltage 100 mV, 1 V, 10 V, 100 V, 750 V ** accuracy in ±(% of measured values )
range |
accuracy 1 year after calibration |
3 – 10 Hz |
0.100 |
10 – 100 Hz |
0.030 |
100 Hz – 1 kHz |
0.010 |
1 – 300 kHz |
0.010 |
obdélníkový signál |
0.010 |
Neelektrické veličiny
Calibration of thermometers contact-less
IR kalibrátor IR-500 – černé těleso
rozsah: 50 to 500°C
rozsah |
accuracy |
< 100°C |
±0.8°C |
100 až 200°C |
±1.6°C |
200 až 500°C |
±2.8°C |
Calibration of contact thermometers
Calibration bath Transcat 7320
Calibration of contact thermometers in the range –20°C to 150 °C. stability:
- ±0,005 °C at –20 °C (ethanol)
- ±0,005 °C at 25 °C (water)
- ±0,007 °C at 150 °C (oil)
Temperature
Multimetr Agilent 34461A
probe |
accuracy 1 year after calibration |
PT100 (DIN/ IEC 751) |
probe accuracy + 0.05 °C |
5 kΩ termistor |
probe accuracy + 0.1 °C |
Humidity
Sensirion EK-H4
sensor |
range |
accuracy of humidity measurement |
accuracy of temperature measurement |
SHT21 |
0 to 100% RH |
± 2% (in the range 20% až 80% RH) |
± 0.3 °C |
SHT71 |
0 to 100% RH |
± 3% (in the range 20% až 80% RH) |
± 0.4 °C |
SHT75 |
0 až 100% RH |
± 1.8% (v rozsahu 10% až 90% RH) |
± 0.3 °C |